Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
Table 6-5: Example of AutoCal Setup
Step
Action
Comment
1.
Press SETUP-ACAL
This button sequence will cause the AUTOCAL menu to be displayed
2.
Press PREV-NEXT
Press PREV-NEXT until SEQ 2 is displayed
3.
Press MODE
Select the MODE menu
3.
Press PREV-NEXT
Press PREV NEXT to scroll to SPAN
4.
Press ENTR
ENTR selects the SPAN MODE
5.
Press SET
Select the SET menu to change the sequence attributes
6.
Press PREV-NEXT
Scroll the SET menu to TIMER ENABLE
7.
Press EDIT
Allows changing the TIMER ENABLE attribute, select ON
8.
Press ENTR
ENTR changes TIMER ENABLE to ON
9.
Press PREV-NEXT
Repeat steps 6-9 for each attribute
10.
Press EXIT
Press the EXIT key to return to upper level menus
6.5 Permeation Tube
NO
2
- nitrogen dioxide - is normally a gas at room temperature and pressure, but can be liquified
at moderate pressures. The permeation tube consists of a small container of NO
2
liquid, with a
small window of PTFE which is permeable to NO
2
, see Figure 6-1. The gas slowly permeates
through the window at a rate in the nanogram/min range. If the tube is kept at constant
temperature, usually about 50
°
C, the device will provide a stable source of NO
2
gas for a year or
more. See below for permeation tube ordering information.
The NO
2
concentration is determined by the permeation tube specific output (ng/min @ 1 slpm
@ 50
o
C), the permeation tube temperature (
o
C) and the air flow across it (slpm). The specific
output in ng/min is a fixed function of the permeation tube and is noted on shipping container.
The ng/min units can be converted to ppb units by the following equation:
⎟
⎠
⎞
⎜
⎝
⎛
⎟⎟
⎠
⎞
⎜⎜
⎝
⎛
=
mole
/
cc
500
,
24
min
/
cc
560
NO
mole
/
gr
46
min)
/
ng
(
output
tube
perm
NO
ppb
2
2
Where:
/mi
perm tube output (ng/min) = The perm tube output in nanograms
nute
ole = The volume of air (at 25 C and 1 atm pressure) that contains 1 mole of air
molecules
46 gr/mole NO
2
= The molecular weight of nitrogen dioxide (NO
2
)
560 cc/min = The flow rate of zero air over the permeation tube
24,500cc/m
6-5
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...