Table 9-4: Diagnostic Mode - Signal I/O (Continued)
No.
Signal
Control
Description
10
BROWNOUT_RST
YES
Brownout reset works in conjunction with
DSP_BROWNOUT. When DSP_BROWNOUT is
set, the CPU sends a signal to reset the display and
clear the DSP_BROWNOUT.
11
CONV_HEATER
YES
Shows the status of the Moly heater. This has the
same function as the LED in the Power Supply
Module.
12 O3GEN_STATUS
YES
Switches
ON/OFF
power to the ozone generator.
13
AUTO_ZERO_VALVE
YES
Switches the AutoZero valve. Use this bit to test the
valve function.
14 PREAMP_RANGE_HI YES
Switches
the preamp(01105) hardware range.
Standard ranges are 2000 and 20,000 ppb. Logic high
= 20,000 ppb; logic low = 2000 ppb. M200A will
reset range to correct value based on user set range
value.
15
ST_RCEL_PRESS
YES
Status Bit - Reaction Cell Pressure alarm
Logic High = pressure out of acceptable range
Logic Low = pressure inside acceptable range
16
ST_ZERO_CAL
YES
Status Bit - Zero Calibration mode
Logic High = M200A in Zero cal mode
Logic Low = Not in Zero cal mode
17
ST_SPAN_CAL
YES
Status Bit - Span Calibration mode
Logic High = M200A in Span cal mode
Logic Low = Not in Span cal mode
18
ST_FLOW_ALARM
YES
Status Bit - Flow alarm
Logic High = Sample/Ozone flow out of spec
Logic Low = Flows within spec
19
ST_TEMP_ALARM
YES
Status Bit - Temperature alarm
Logic High = Rxcell, Moly, IZS, Box temps out of
spec
Logic Low = Temps within spec
20
ST_DIAG_MODE
YES
Status Bit - In Diagnostic mode
Logic High = M200A in Diagnostic mode
Logic Low = Not in Diag mode
(table continued)
9-14
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...