Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
Table 9-6: Test Channel Readings (Continued)
TEST Channel
Minimum*
Maximum*
Description
IZS TEMP
0
o
C
70
o
C
The IZS temperature is set to 50
o
C. At
the setpoint, a typical reading is
3600 mV.
CONV TEMP
0
o
C
1000
o
C
The Converter temperature is 315
o
C. At
the setpoint, a typical voltage is
3150 mV.
PMT TEMP
0
o
C
70
o
C
The PMT temperature is unique in that
the voltage is inverse to the temperature.
A typical reading for 8
o
C would be
4200 mV.
CHASSIS
TEMP
0
o
C
70
o
C
The Chassis (Box) temperature is
variable due to variable ambient air
temperature. The Box temp generally
runs about 5
o
C above the surrounding
air temp. Thus in a 25
o
C room, the Box
temp would be about 30
o
C and have a
TEST channel voltage of about
2000 mV.
DCPS
VOLTAGE
0 mV
5000 mV
The DCPS is a composite of several DC
power supply voltages in the instrument.
It has been arbitrarily set at 2500 mV,
which is typical.
HVPS
VOLTAGE
0 V
5000 V
The HVPS voltage is a scaled up
reading of the programming voltage
going to the HVPS. Zero to 1000 mV
corresponds 0-1000 VDC for the HVPS,
which is the maximum voltage possible.
A typical reading would be 700 mV
corresponding to 700 VDC for the
HVPS.
* Minimum and Maximum readings depend on the DAC 3 switch settings of the V/F board. For
the standard
±
5 VDC range, minimum corresponds to 0 VDC and maximum corresponds to
5 VDC.
9-22
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...