Analyzer Instruction Manual, 02246, Rev. F1
Table 2-1: Final Test and Calibration Values
Test Values
Observed
Value
Units
Nominal Range
Reference Section
RANGE PPB
50-20,000 5.3.4
STABILITY
(Zero Gas)
PPB
< 0.2
SAMP FLW
cc/min
500 ± 50
OZONE FL
cc/min
60 - 90
PMT
mV
0-5000
AZERO
mV
-20 - 150
HVPS
V
450 - 900 constant
DCPS
mV
2500 ± 200
RCELL TEMP
o
C 50
±1
BOX TEMP
o
C 8-48
PMT TEMP
o
C
7 ± 1
IZS TEMP
o
C
50 ± 0.4
MOLY TEMP
o
C
315 ± 5
RCEL PRES
IN-Hg-A
4 - 10 constant
SAMP PRES
IN-Hg-A
25 - 30 constant
Electric Test & Optic Test
Electric Test
PMT Volts
mV
2000 ± 500
NO Conc
PPB
1000 ± 250
NO
x
Conc
PPB
1000 ± 250
Optic Test
PMT Volts
mV
2000 ± 1000
NO Conc
PPB
1000 ± 500
NO
x
Conc
PPB
1000 ± 500
(table continued)
2-12
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...