Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
4.
Calculate the resulting NO
2
concentrations as follows:
F
]
NO
[
*
F
+
]
[NO
-
]
[NO
=
]
NO
[
Where
[NO]
T
IMP
2
NO
REM
ORIG
OUT
2
Equation 7-10
:
ncentration before the GPT ozone is turned on, and [NO]
REM
is the NO
rem
5.
ause is
that the converter needs replacing. See Section 7.8.6 on Moly converter efficiency.
7.8
e
determined from the data collected in 7.8.5.3.
1.
Calculate the concentration of NO
2
converted as:
REM
x
ORIG
x
OUT
2
NV
Equation 7-11
:
ntration before the GPT ozone is turned on, and [NO
x
]
REM
is the
NO
2.
ould be a straight line within the
± 2
3.
iciency. The value should be between .96
and
4.
key in the slope from step 3 followed by ENTR. Press EXIT to return to the SAMPLE menu.
ORIG
is the NO co
aining after GPT.
Plot the NO
2
concentration output by the instrument on the y-axis against the generated
NO
2
[NO
2
]
OUT
on the x-axis. The plot should be a straight line within the ± 2% linearity
criteria given for the NO
x
and NO channels. If the plot is not linear the most likely c
.6 Moly Converter Efficiency
The moly efficiency should be 96 to 102% efficient. If it is outside these limits it should b
replaced. The converter efficiency can be
For each NO
2
concentration generated:
[
CO
2
Where
[NO
)
]
NO
[
-
]
NO
([
-
]
NO
[
=
]
NO
x
]
ORIG
is the NO
x
conce
x
remaining after GPT.
Plot the [NO
2
]
CONV
concentration output by the instrument on the y-axis against the
concentration generated [NO
2
]
OUT
on the x-axis. The plot sh
% linearity criteria given for the NO
x
and NO channels.
Determine the best straight line fit of the plot either by inspection or least squares. The
slope of the resulting straight line is the moly eff
1.02. If not, the moly needs to be replaced.
If you want the M200A to automatically compensate (Section 7.8.6.1) for converter
efficiency, enter the efficiency value in the front panel by CAL-CONC-MOLY-SET, then
7-25
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...