Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
7.2 Manual Zero/Span Check or Calibration with
Zero/Span Valves Option
The Zero/Span valve option can be operated from the front panel keyboard. In the Zero/Span
valve option the zero and span gas come into the valves through ports on the rear panel of the
instrument.
Table 7-5: Manual Zero Calibration Procedure - Z/S Valves
Step Number
Action
Comment
1.
Press CALZ
The analyzer enters the zero calibrate mode. This switches the
sample/cal and zero/span valves to allow zero gas to come in
through the zero gas inlet port in the rear panel.
2.
Wait 10 min
Wait for reading to stabilize at zero value.
3.
Press ZERO
If you change your mind after pressing ZERO, you can still
press EXIT here without zeroing the instrument.
4.
Press ENTR
Pressing ENTR actually changes the calculation equations,
forcing the reading to zero.
5.
Press EXIT
M200A returns to sample mode. Immediately after
calibration, readings do not go into the DAS averages.
Refer to Table 7-3 to enter expected NO and NO
x
values.
Table 7-6: Manual Span Calibration Procedure - Z/S Valves
Step Number
Action
Comment
1.
Press CALS
The M200A enters the calibrate mode from sample mode.
This operates the sample/cal and zero/span valves to allow
span gas to come in through the cal gas inlet port in the rear
panel or optional permeation tube.
2.
Wait 10 min
Wait for reading to stabilize at span value.
3.
Press SPAN
If you change your mind after pressing SPAN, you can still
press EXIT here without spanning the instrument.
4.
Press ENTR
Pressing ENTR actually changes the calculation equations.
5.
Press EXIT
M200A returns to sampling. Immediately after calibration,
data is not added to the DAS averages.
7-6
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...