Table 9-4: Diagnostic Mode - Signal I/O (Continued)
No.
Signal
Control
Description
21
ST_POWER_OK
YES
Status Bit - Power OK
Logic High = Instrument power is on
Logic Low = Instrument power is off
22
ST_SYSTEM_OK
YES
Status Bit - System OK
Logic High = No instrument warnings present
Logic Low = 1 or more alarms present
23
ST_HIGH_RANGE
YES
Status Bit - Autorange High Range
Logic High = M200A in high range of autorange
mode
Logic Low = M200A in low range of autorange
mode
24
PMT_SIGNAL
NO
Current PMT voltage. Same as PMT voltage in TEST
menu. Bi-polar, typically in 0-5000 mV range. A
constant value of 5000 mV indicates offscale.
25
SAMPLE_PRESs
NO
Sample pressure in mV. Typical sea level value =
4300 mV for 29.9" Hg-A.
26
RCELL_TEMP
NO
Reaction Cell temperature. Typically 3500 mV for
50
°
C.
27 BOX_TEMP
NO
Box Temperature. Typically 1800 mV for 25
°
C
28
IZS_TEMP
NO
IZS permeation tube oven temp. Typically 3500 mV
for 50
°
C.
29
PMT_TEMP
NO
PMT cold block temperature. Typically 3600 mV for
10
°
C.
30
DCPS_VOLTAGE
NO
DC power supply composite voltage output.
Typically 2500 mV.
31
RCELL_PRESS
NO
Reaction Cell Pressure in mV. Typically 1270 mV
for 5" Hg-A at sea level. Is an absolute pressure so
higher values means higher absolute pressures.
32
OZONE_FLOW
NO
Ozone flowmeter voltage. Typically 2000 mV at
80 cc/min.
33
CONV_TEMP
NO
Molybdenum Converter temp. Typically 3150 mV at
315
°
C.
34
HVPS_VOLTAGE
NO
HVPS programming voltage. Output of HVPS is
1000x value present. 700 mV = 700 VDC output.
(table continued)
9-15
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...