Analyzer Instruction Manual, 02246, Rev. F1
Table 9-2: Front Panel Warning Messages
Message
Description
SYSTEM RESET
Analyzer was reset/powered on. This warning occurs every
time the instrument is powered up, as in a power failure. It can
also occur if the RAM or EEPROM is reset.
RAM INITIALIZED
RAM was erased. The RAM contains the DAS averages which
get erased when the RAM is initialized. It also contains
temporary data used by the M200A to calculate concentrations.
No setup variables are stored in the RAM.
SAMPLE FLOW WARN
The calculated sample flow is outside the hi/low limits. Since
the flow is calculated, it probably means the pressure has
gotten too low. This can be caused by a plugged sample inlet,
or water in ozone line into the reaction cell.
OZONE FLOW WARNING
Ozone flow out of spec. warnings occur most often because of
loss of vacuum, which causes the ozone flow to go to zero.
They also can occur due to a flow sensor failure.
RxCELL PRESS WARNING
Vacuum out of spec. warnings are caused by leaks, pump
failure or disconnected pump.
BOX TEMP WARNING
Box temp. out of spec. Instrument fan failure, enclosure
temperature failure. Operation of the M200A in a too warm or
cold environment will cause degradation of data quality and
shorten the life of the instrument.
RCELL TEMP WARNING
Reaction cell temp. out of spec. The warning message is most
often present during initial warm-up or if the connector to the
heaters is not plugged in after dis-assembly. It has also
occurred if the thermistor is not in position in the reaction cell.
IZS TEMP WARNING
IZS temp. out of spec. The warning message is most often
present during initial warm-up or if the connector to the
heaters is not plugged in after dis-assembly. Also check the
position of the thermistor in permeation tube heater block.
Check to see if there is a good seal at the top of the heater
block.
(table continued)
9-9
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...