Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
7.6
Calibrate on NO
2
Permeation Tube
If the M200A is equipped with the IZS option, is possible to span the instrument on the
permeation tube. If this feature is enabled, the instrument internal valves are set so that gas is
routed through the permeation tube, then the moly converter, thus producing NO gas. The
software is programmed to use this known NO gas concentration to span both the NO and NO
x
channels of the instrument. Thus, if CAL-ON-NO2 is enabled, the CALS button now will span
both the NO and NO
x
channels of the instrument.
CAUTION
This feature has some unexpected side effects.
tration
us
tration of NO gas in the NO and NOx channels, and
reports a NO concentration of zero.
disabled so that NO
2
concentrations are reported correctly. See Table 7-8 to enable this feature.
If CAL-ON-NO2 is enabled and you do a SPAN CHECK or SPAN CAL, the NO
2
concen
will be reported as ZERO! This is because all of the NO
2
gas is being routed through the
molybdenum converter during both the NO and NO
x
portions of the measurement cycle. Th
the instrument sees the SAME concen
2
As long as CAL-ON-NO2 is enabled the instrument will never report any NO
2
during span
calibration. This feature should be enabled only when a calibration is desired, otherwise leave it
NOTE
Note that this method of calibration is NOT approved by the USEPA.
Permeation tubes are known to change as temperatures are cycled,
and as they age. Also, this calibration relies on the moly converter
to be working properly. (Converter efficiency can be compensated
by enabling the Converter Efficiency option.) It is recommended that
if CAL-ON-NO2 is used, the instrument calibration be checked
frequently by independent means to assure accurate data.
7-9
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...