Table 8-2: Preventative Maintenance Schedule
Item
Jan
Feb
Mar Apr
May Jun
Jul
Aug
Sep
Oct
Nov
Dec
Maintenance
Interval
TEST
Functions
6 - 12 month
intervals
Zero/Span
Calibration
Annually or after
repairs
Zero/Span Checks
y
Dail
Particulate
Filter
Weekly
as
needed
Ozone Scrubber on
Pump Pack
Replace charcoal
every 3 months
Filter for Ozone
Scrubber
Replace every 12
months
Filter for PermaPure
Drier
Replace every 12
months
Zero Air Canister
Refill every 3
months
Reaction Cell
Window
Clean annually or as
necessary
Ozone
Flow
Check
every
year
Sample
Flow
Check
every
year
Moly
Converter
Check efficiency
every 6 months
Pneumatic
Lines
Examine every 12
months, clean if
necessary
Factory
Calibration
Calibrate each year
or after repairs
Leak
Check
Check
every
year
O-rings
Replace if seal is
broken
Teledy
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...