Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
Because low flows are difficult to control and measure, it is often advantageous to set a higher
total flow than needed. In this example, we will let F
T
= 2750 cm3/min
2.
Determine the highest NO concentration, [NO]
OUT
, required at the output manifold, using
Equation 7-2:
[NO]
OUT
= 0.5 ppm (90/100) = 0.45 ppm
3.
Calculate the NO flow (F
NO
) required to generate the NO concentration [NO]
OUT
, using
Equation 7-3:
min
/
cm
24.5
ppm
5
.
50
min
/
cm
2750
ppm
45
.
0
=
F
3
3
NO
=
×
4.
Calculate the required flow rate through ozone generator (F
O
) using Equation 7-4:
O
3
3
3
F =
50.5 ppm x 24.5 cm /
x 180 cm
2.75 ppm -
- 24.5 cm /
min
min
min
= 80984
cm /
- 24.5 cm /
= 260 cm /
6
2
3
3
min
min
min
5.
Verify that the residence time (t
R
) in the reaction chamber is <2 min using Equation 7-5:
min
63
.
0
min
/
cm
260
cm
180
=
t
3
3
R
=
min
/
cm
5
.
24
3
+
6.
Verify the dynamic parameter specification (P
R
) of the calibrator reaction chamber using
Equation 7-6:
min
ppm
75
.
2
min
/
cm
5
.
24
min
/
cm
260
min
/
cm
5
.
24
min
/
cm
260
ppm
5
.
50
=
3
3
3
3
−
=
+
×
+
×
cm
180
min
/
cm
24
P
3
3
R
7.
Calculate the diluent air flow (F
D
) required at the mixing chamber, using Equation 7-7:
F
D
= 2750 cm
3
/min - 260 cm
3
/min - 24.5 cm
3
/min = 2465.5 cm
3
/min
7-19
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...