
Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
7.8.5.1 Zero Calibration Procedure
Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the expected zero
value. The following Table 7-12 details the zero calibration procedure.
Table 7-12: Zero Calibration Procedure
Step Number
Action
Comment
1.
Press CAL
The M200A enters the calibrate mode from sample mode.
NOTE:
The analyzer does not operate the zero/span valves in this mode,
the zero gas enters through the sample port.
2.
Wait 10 min
Wait for reading to stabilize at the zero value.
3.
Press ZERO
If you change your mind after pressing ZERO, you can still
press EXIT here without zeroing the instrument.
4.
Press ENTR
Pressing ENTR actually changes the calculation equations.
5.
Press EXIT
M200A returns to the SAMPLE mode.
7.8.5.2 NO/NO
x
Calibration Procedure
Adjust the NO concentration to approximately 80% of the URL of the NO channel. The expected
NO and NO
x
span concentrations can be determined by measuring the cylinder and diluent flows
and computing the resulting concentrations. If there is any NO
2
impurity in the NO standard gas
it should be taken into account when the NO
x
concentration is entered during the NO/NO
x
channel calibration. This is done by ADDING the impurity concentration to the NO
concentration to get the NO
x
concentration for calibration. Calculate the exact NO and NO
x
concentrations as follows:
F
]
[NO
x
F
=
]
[NO
T
STD
NO
OUT
Equation 7-8
Enter the respective concentrations using the
procedure in Table 7-13. The expected span concentrations need not be re-entered each time a
calibration is performed unless they are changed.
Enter the expected NO
x
and NO span gas concentrations:
7-22
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...