Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
Table 7-8: Enabling CAL-ON-NO2
Step Number
Action
Comment
1. Press
SETUP-
MORE-VARS
Causes the M200A go to the VARS(Variables) menu
2.
Enter the VARS
password, then
ENTR
Password value is 8 1 8
3.
Scroll the VARS
menu to select the
CAL-ON-NO2
variable
By pressing NEXT or PREV
4.
Press EDIT
Once you have entered EDIT, set the value to ON
5.
Press ENTR
ENTR will enable the CAL-ON-NO2 feature
6.
Press EXIT
Each EXIT press returns the M200A to the next higher menu
level. The highest level is the SAMPLE mode
Once CAL-ON-NO2 is enabled the procedure for zeroing and spanning the instrument is
identical to that given in Section 7.2 for Zero/Span Valves.
7-10