6.
Navigate to
SNMP Trap Enable
and press the right arrow key to select
On
.
This setting turns the traps that are sent out from the instrument through
SNMP on and off.
7.
Press the
CONFIG
button on the front panel to close the menu display.
Incoming inspection
The incoming inspection procedures are basic optional procedures to check the
functionality of your instrument. These procedures require no external
equipment.
For a more robust inspection, see the performance verification procedures in the
WFM2300 and WFM2200A Specifications and Performance Verification
Technical Reference
.
Basic turn on and self test
1.
Connect power to the instrument. See
2.
Turn the instrument on. See
3.
After about 30 seconds, the power-on diagnostic page should appear on the
screen.
4.
Verify that all self tests pass. Any failures are shown in red.
NOTE.
Although the results of the power-on diagnostics are erased from the
screen during the power-on process, you can view the results by selecting or
by viewing them on the instrument Web page.
5.
After the diagnostics are finished, the instrument state is restored. When the
progress indicator in the status bar is finished, the instrument has finished
initializing.
Front panel test
1.
Reset the instrument settings to the Factory Presets:
a.
Press and hold the
PRESET
button to open the Preset menu.
b.
Select
Recall Preset
>
Recall Factory Preset
, and then press the
SEL
button. Wait for the process to complete as indicated by the progress
indicator.
2.
Press the
HELP
button to display the instrument help.
3.
Press each of the front panel buttons, one at a time, except for the
Power
and
HELP
buttons.
Getting started
24
WFM2300 and WFM2200A Waveform Monitors User Manual
Summary of Contents for WFM2200A
Page 2: ......
Page 6: ......
Page 16: ...Table of Contents x WFM2300 and WFM2200A Waveform Monitors User Manual ...
Page 200: ...Functions 178 WFM2300 and WFM2200A Waveform Monitors User Manual ...
Page 208: ...Software upgrades 186 WFM2300 and WFM2200A Waveform Monitors User Manual ...
Page 242: ...ARIB content displays 220 WFM2300 and WFM2200A Waveform Monitors User Manual ...
Page 252: ...Audio monitoring 230 WFM2300 and WFM2200A Waveform Monitors User Manual ...
Page 270: ...Application examples 248 WFM2300 and WFM2200A Waveform Monitors User Manual ...