:TRIGger Commands
2- 560
WCA230A & WCA280A Programmer Manual
:TRIGger[:SEQuence]:MPOSition? (Query Only)
Queries the trigger occurence point in one block data acquired on the memory
when measurement results are obtained with the :FETCh or :READ commands.
:TRIGger[:SEQuence]:MPOSition? <value>
<value>::=<NR1>
specifies the block number. Zero indicates the latest block.
Range: --2285 to 0 (standard) or --9142 to 0 (option 02)
<NR1>
represents the trigger occurence point. The returned value depends on
whether a trigger occurred or not, as shown in the table below.
Trigger occurrence
Returned value
1
Trigger occurred
--1024 to (block size)
×
1024 --1
No trigger occurred
(block size)
×
1024
1
The block size is set with [:SENSe]:BSIZe.
A minus value indicates that the trigger occurred before the block data acquisi-
tion.
If you send :TRIGger[:SEQuence]:MPOSition? MINimum | MAXimum when
the measurement is not performed, “Execution error” (--200) is returned.
NOTE
. When you select PNEGative or NPOSitive with the :TRIGger[:SEQuence]
:SLOPe command or IQFRequency with the :TRIGger[:SEQuence]:SOURce
command, the returned value is the same as the :TRIGger[:SEQuence]:OPOSi-
tion? query because the analyzer cannot determine the trigger occurrence point.
SARTIME, SAZRTIME, all Demod modes, all Time modes
:TRIGger:SEQuence:MPOSition? -15
might return 123, indicating that the trigger occurred at the 123th data point in
the block #--15.
[:SENSe]:BSIZe, :TRIGger[:SEQuence]:OPOSition?,
:TRIGger[:SEQuence]:SLOPe, :TRIGger[:SEQuence]:SOURce
Syntax
Arguments
Returns
Measurement Modes
Examples
Related Commands
Summary of Contents for WCA230A
Page 13: ...Getting Started...
Page 14: ......
Page 24: ...Getting Started 1 10 WCA230A WCA280A Programmer Manual...
Page 25: ...Syntax and Commands...
Page 26: ......
Page 92: ...CALCulate Commands 2 66 WCA230A WCA280A Programmer Manual...
Page 100: ...CALibration Commands 2 74 WCA230A WCA280A Programmer Manual...
Page 114: ...CONFigure Commands 2 88 WCA230A WCA280A Programmer Manual...
Page 270: ...DISPlay Commands 2 244 WCA230A WCA280A Programmer Manual...
Page 320: ...FETCh Commands 2 294 WCA230A WCA280A Programmer Manual...
Page 326: ...HCOPy Commands 2 300 WCA230A WCA280A Programmer Manual...
Page 330: ...INITiate Commands 2 304 WCA230A WCA280A Programmer Manual...
Page 350: ...MMEMory Commands 2 324 WCA230A WCA280A Programmer Manual...
Page 356: ...PROGram Commands 2 330 WCA230A WCA280A Programmer Manual...
Page 396: ...READ Commands 2 370 WCA230A WCA280A Programmer Manual...
Page 554: ...SENSe Commands 2 528 WCA230A WCA280A Programmer Manual...
Page 562: ...STATus Commands 2 536 WCA230A WCA280A Programmer Manual...
Page 572: ...SYSTem Commands 2 546 WCA230A WCA280A Programmer Manual...
Page 578: ...TRACe Commands 2 552 WCA230A WCA280A Programmer Manual...
Page 594: ...TRIGger Commands 2 568 WCA230A WCA280A Programmer Manual...
Page 598: ...Retrieving Response Message 2 572 WCA230A WCA280A Programmer Manual...
Page 599: ...Status and Events...
Page 600: ......
Page 616: ...Status and Events 3 16 WCA230A WCA280A Programmer Manual...
Page 623: ...Programming Examples...
Page 624: ......
Page 641: ...Appendices...
Page 642: ......
Page 660: ...Appendix C Factory Initialization Settings C 12 WCA230A WCA280A Programmer Manual...
Page 668: ...Appendix E SCPI Conformance Information E 4 WCA230A WCA280A Programmer Manual...
Page 669: ...Glossary and Index...
Page 670: ......
Page 672: ...Glossary Glossary 2 WCA230A WCA280A Programmer Manual...
Page 676: ...Index Index 4 WCA230A WCA280A Programmer Manual...
Page 677: ......
Page 678: ......