The SESR is made up of 8 bits. Each bit records the occurrence of a different
type of event, as shown in Figure 3--3 and Table 3--2. The contents of this
register are returned when the *ESR? query is used.
PON
CME EXE DDE QYE
OPC
7
6
5
4
3
2
1
0
—
—
Figure 3- 3: The Standard Event Status Register (SESR)
Table 3- 2: SESR bit functions
Bit
Function
7
Power On (PON). Indicates that the power to the instrument is on.
6
Not used.
5
Command Error (CME). Indicates that a command error has occurred while
parsing by the command parser was in progress.
4
Execution Error (EXE). Indicates that an error occurred during the execution of
a command. Execution errors occur for one of the following reasons:
H
When a value designated in the argument is outside the allowable range
of the instrument, or is in conflict with the capabilities of the instrument
H
When the command could not be executed properly because the
conditions for execution differed from those essentially required
3
Device-Dependent Error (DDE). An instrument error has been detected.
2
Query Error (QYE). Indicates that a query error has been detected by the
output queue controller. Query errors occur for one of the following reasons:
H
An attempt was made to retrieve messages from the output queue,
despite the fact that the output queue is empty or in pending status.
H
The output queue messages have been cleared despite the fact that they
have not been retrieved.
1
Not used.
0
Operation Complete (OPC). This bit is set with the results of the execution of
the *OPC command. It indicates that all pending operations have been
completed.
Standard Event Status
Register (SESR)
Summary of Contents for WCA230A
Page 13: ...Getting Started...
Page 14: ......
Page 24: ...Getting Started 1 10 WCA230A WCA280A Programmer Manual...
Page 25: ...Syntax and Commands...
Page 26: ......
Page 92: ...CALCulate Commands 2 66 WCA230A WCA280A Programmer Manual...
Page 100: ...CALibration Commands 2 74 WCA230A WCA280A Programmer Manual...
Page 114: ...CONFigure Commands 2 88 WCA230A WCA280A Programmer Manual...
Page 270: ...DISPlay Commands 2 244 WCA230A WCA280A Programmer Manual...
Page 320: ...FETCh Commands 2 294 WCA230A WCA280A Programmer Manual...
Page 326: ...HCOPy Commands 2 300 WCA230A WCA280A Programmer Manual...
Page 330: ...INITiate Commands 2 304 WCA230A WCA280A Programmer Manual...
Page 350: ...MMEMory Commands 2 324 WCA230A WCA280A Programmer Manual...
Page 356: ...PROGram Commands 2 330 WCA230A WCA280A Programmer Manual...
Page 396: ...READ Commands 2 370 WCA230A WCA280A Programmer Manual...
Page 554: ...SENSe Commands 2 528 WCA230A WCA280A Programmer Manual...
Page 562: ...STATus Commands 2 536 WCA230A WCA280A Programmer Manual...
Page 572: ...SYSTem Commands 2 546 WCA230A WCA280A Programmer Manual...
Page 578: ...TRACe Commands 2 552 WCA230A WCA280A Programmer Manual...
Page 594: ...TRIGger Commands 2 568 WCA230A WCA280A Programmer Manual...
Page 598: ...Retrieving Response Message 2 572 WCA230A WCA280A Programmer Manual...
Page 599: ...Status and Events...
Page 600: ......
Page 616: ...Status and Events 3 16 WCA230A WCA280A Programmer Manual...
Page 623: ...Programming Examples...
Page 624: ......
Page 641: ...Appendices...
Page 642: ......
Page 660: ...Appendix C Factory Initialization Settings C 12 WCA230A WCA280A Programmer Manual...
Page 668: ...Appendix E SCPI Conformance Information E 4 WCA230A WCA280A Programmer Manual...
Page 669: ...Glossary and Index...
Page 670: ......
Page 672: ...Glossary Glossary 2 WCA230A WCA280A Programmer Manual...
Page 676: ...Index Index 4 WCA230A WCA280A Programmer Manual...
Page 677: ......
Page 678: ......