:READ Commands
2- 346
WCA230A & WCA280A Programmer Manual
:READ:OVIew? (Query Only)
Obtains the minimum and maximum values for each 1024-point segment of
waveform data displayed on the overview in the Demod (modulation analysis)
and the Time (time analysis) modes.
NOTE
. The :CONFigure:OVIew command must be run to turn the measurement
off before the :READ:OVIew command is executed.
:READ:OVIew?
#<Num_digit><Num_byte><MinData(1)><MaxData(1)>...
<MinData(n)><MaxData(n)>
Where
<Num_digit>
is the number of digits in <Num_byte>.
<Num_byte>
is the number of bytes of the data that follow.
<MinData(n)>
is the minimum data in dBm for each 1024 data point segment.
4-byte little endian floating-point format specified in IEEE 488.2
<MaxData(n)>
is the maximum data in dBm for each 1024 data point segment.
4-byte little endian floating-point format specified in IEEE 488.2
n: Max 16000 (standard) / 64000 (Option 02)
All Demod modes, all Time modes
:READ:OVIew?
might return
#510240xxx...
(10240-byte data) representing the minimum and
the maximum values of waveform displayed on the overview.
:CONFigure:OVIew, :INSTrument[:SELect]
Syntax
Returns
Measurement Modes
Examples
Related Commands
Summary of Contents for WCA230A
Page 13: ...Getting Started...
Page 14: ......
Page 24: ...Getting Started 1 10 WCA230A WCA280A Programmer Manual...
Page 25: ...Syntax and Commands...
Page 26: ......
Page 92: ...CALCulate Commands 2 66 WCA230A WCA280A Programmer Manual...
Page 100: ...CALibration Commands 2 74 WCA230A WCA280A Programmer Manual...
Page 114: ...CONFigure Commands 2 88 WCA230A WCA280A Programmer Manual...
Page 270: ...DISPlay Commands 2 244 WCA230A WCA280A Programmer Manual...
Page 320: ...FETCh Commands 2 294 WCA230A WCA280A Programmer Manual...
Page 326: ...HCOPy Commands 2 300 WCA230A WCA280A Programmer Manual...
Page 330: ...INITiate Commands 2 304 WCA230A WCA280A Programmer Manual...
Page 350: ...MMEMory Commands 2 324 WCA230A WCA280A Programmer Manual...
Page 356: ...PROGram Commands 2 330 WCA230A WCA280A Programmer Manual...
Page 396: ...READ Commands 2 370 WCA230A WCA280A Programmer Manual...
Page 554: ...SENSe Commands 2 528 WCA230A WCA280A Programmer Manual...
Page 562: ...STATus Commands 2 536 WCA230A WCA280A Programmer Manual...
Page 572: ...SYSTem Commands 2 546 WCA230A WCA280A Programmer Manual...
Page 578: ...TRACe Commands 2 552 WCA230A WCA280A Programmer Manual...
Page 594: ...TRIGger Commands 2 568 WCA230A WCA280A Programmer Manual...
Page 598: ...Retrieving Response Message 2 572 WCA230A WCA280A Programmer Manual...
Page 599: ...Status and Events...
Page 600: ......
Page 616: ...Status and Events 3 16 WCA230A WCA280A Programmer Manual...
Page 623: ...Programming Examples...
Page 624: ......
Page 641: ...Appendices...
Page 642: ......
Page 660: ...Appendix C Factory Initialization Settings C 12 WCA230A WCA280A Programmer Manual...
Page 668: ...Appendix E SCPI Conformance Information E 4 WCA230A WCA280A Programmer Manual...
Page 669: ...Glossary and Index...
Page 670: ......
Page 672: ...Glossary Glossary 2 WCA230A WCA280A Programmer Manual...
Page 676: ...Index Index 4 WCA230A WCA280A Programmer Manual...
Page 677: ......
Page 678: ......