Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–81
c.
Check Sync Duty Cycle:
H
Connect the pulse generator through a 50
W
cable and a 50
W
terminator to the oscilloscope CH1 input. See Figure 1–38.
Pulse
Generator
Digitizing Oscilloscope
50
cable
50
terminator
Output
Figure 1–38: Setup for sync duty cycle test
H
Turn the pulse generator OUTPUT (VOLTS) control until the signal
on the oscilloscope shows a one division negative going pulse. See
Figure 1–39.
NOTE
. You may need to adjust the trigger level control to obtain a stable trigger.