![Tektronix TDS 500C Performance Verification Manual Download Page 127](http://html1.mh-extra.com/html/tektronix/tds-500c/tds-500c_performance-verification-manual_1079136127.webp)
Nominal Traits
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
2–11
Table 2–4: Nominal traits — Time base system (cont.)
Name
Description
Record Length Selection
TDS 500C & TDS 700C:
500, 1,000, 2,500, 5,000, 15,000 and 50,000 samples
In addition to the record lengths previously listed, the following record lengths are
available with the following options:
TDS 520C & TDS 724 with option 1M:
1-channel:
up to 250,000 samples
2-channels:
up to 130,000 samples
TDS 520C & TDS 724 with option 2M:
1-channel:
up to 4,000,000 samples
2-channels:
up to 2,000,000 samples
TDS 540C, TDS 754C & TDS 784C with option 1M:
1-channel:
up to 500,000 samples
2-channels:
up to 250,000 samples
3 or 4-channels
up to 130,000 samples
TDS 540C, TDS 754C & TDS 784C with option 2M:
1-channel:
up to 8,000,000 samples
2-channels:
up to 4,000,000 samples
3 or 4-channels
up to 2,000,000 samples
1
The range of real-time rates, expressed in samples/second, at which a digitizer samples signals at its inputs and stores
the samples in memory to produce a record of time-sequential samples.
2
The range of waveform rates for interpolated (or equivalent-time on the TDS 700C) waveform records.
3
The Waveform Rate (WR) is the equivalent sample rate of a waveform record. For a waveform record acquired by
real-time sampling of a single acquisition, the waveform rate is the same as the real-time sample rate; for a waveform
created by interpolation of real-time samples from a single acquisition or, on applicable products, the equivalent-time
sampling of multiple acquisitions, the waveform rate created is faster than the real time sample rate. For all these cases,
the waveform rate is 1/(Waveform Interval) for the waveform record, where the waveform interval (WI) is the time between
the samples in the waveform record.
Table 2–5: Nominal traits — Triggering system
Name
Description
Range, Delayed Trigger Time Delay
16 ns to 250 s
Range, Events Delay
TDS 600B; 2 to 10,000,000
TDS 500C/700C: 1 to 10,000,000
Range (Time) for Pulse-Glitch,
Pulse-Width, Time-Qualified Runt, Timeout,
or Slew Rate Trigger, Delta Time
1 ns to 1 s
Summary of Contents for TDS 500C
Page 4: ......
Page 15: ...Performance Verification Procedures ...
Page 16: ......
Page 115: ...Specifications ...
Page 116: ......
Page 132: ...Nominal Traits 2 16 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Page 147: ......
Page 148: ......