Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–73
Positive pulse
Negative pulses
Figure 1–30: Triggered signal range test – 75 mV
p. Disconnect all test equipment (TSG121) from the digitizing oscillo-
scope.
5. Check 60 Hz Rejection.
a. Set up oscilloscope for 60 Hz Rejection Test:
H
Use the keypad to set the Ch1 Fine Scale to 282 mV (press 282,
SHIFT m, then ENTER).
H
Press WAVEFORM OFF.
H
Press CH2.
H
Press VERTICAL MENU.
H
Use the keypad set the fine scale to 2 V (press 2, then ENTER).
H
Press HORIZONTAL MENU.
H
Use the keypad to set the horizontal scale (/div) to 5 ms (press 5,
SHIFT, m, then ENTER).
b. Set up 60 Hz signal generator:
H
Connect the output of the signal generator to the CH2 input through
a 50
W
cable. See Figure 1–31.