Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–63
c.
Modify the initialized front-panel control settings:
H
Set the horizontal SCALE to 200
s.
H
Press SET LEVEL TO 50%.
H
Use the vertical POSITION knob to center the display on screen.
H
Press SHIFT; then press ACQUIRE MENU.
H
Press the main-menu button Mode; then press the side-menu button
Average.
H
Select 128 averages with the keypad or the general purpose knob.
2. Confirm that the Probe Compensator signal is within limits for frequency:
a. Measure the frequency of the probe compensation signal:
H
Press MEASURE; then press the main-menu button Select
Measrmnt for Ch1.
H
Repeatedly press the side-menu button –more– until Frequency
appears in the side menu (its icon is shown at the left). Press the
side-menu button Frequency.
b. Check against limits:
H
CHECK that the CH 1 Freq readout is within 950 Hz to 1.050 kHz,
inclusive. See Figure 1–23.
H
Enter frequency on test record.
H
Press MEASURE; then press the main-menu button Remove
Measrmnt. Press the side-menu Measurement 1.
Summary of Contents for TDS 500C
Page 4: ......
Page 15: ...Performance Verification Procedures ...
Page 16: ......
Page 115: ...Specifications ...
Page 116: ......
Page 132: ...Nominal Traits 2 16 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Page 147: ......
Page 148: ......