Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–39
b. Modify the initialized front-panel control settings:
H
Do not adjust the vertical position of any channel during this
procedure.
H
Set the horizontal SCALE to 500 ps.
H
Press SHIFT; then press ACQUIRE MENU.
H
Press the main-menu button Mode, and then press the side-menu
button Average 16.
c.
Hook up the test-signal source:
H
Connect the sine wave output of a sine wave generator (item 19 or,
optionally, 25) to a 50
W
precision coaxial cable followed by a 50
W
termination, and a dual-input coupler. See Figure 1–10.
H
Connect the coupler to both CH 1 and CH 2. See Figure 1–10.
Sine Wave
Generator
Dual input
coupler
Digitizing Oscilloscope
50
terminator
Output
Figure 1–10: Initial test hookup
2. Confirm all four channels are within limits for channel delay:
a. Set up the generator: Set the generator frequency to 250 MHz and the
amplitude for about six divisions in CH 1.
Hint: As you are adjusting the generator amplitude, push SET LEVEL
TO 50% frequently to speed up the updating of the waveform amplitude
on screen.
Summary of Contents for TDS 500C
Page 4: ......
Page 15: ...Performance Verification Procedures ...
Page 16: ......
Page 115: ...Specifications ...
Page 116: ......
Page 132: ...Nominal Traits 2 16 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Page 147: ......
Page 148: ......