![Tektronix TDS 500C Performance Verification Manual Download Page 76](http://html1.mh-extra.com/html/tektronix/tds-500c/tds-500c_performance-verification-manual_1079136076.webp)
Performance Tests
1–60
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
Check output
Figure 1–21: Measurement of main trigger out limits
d. Check Delayed Trigger output against limits: See Figure 1–21.
H
Move the precision 50
W
cable from the rear-panel Main Trigger
Output BNC to the rear-panel Delayed Trigger Output BNC.
H
CHECK that the Ch2 High readout is
≥
1.0 volt and that the Ch2
Low readout
≤
0.25 volts.
H
Enter high and low voltages on test record.
H
Press the side-menu button
to select the 1 M
W
setting.
H
Press CLEAR MENU.
H
CHECK that the Ch2 High readout is
≥
2.5 volts and that the Ch2
Low readout is
≤
0.7 volts.
H
Enter high and low voltages on test record.
3. Confirm CH 3 (Ax1 on some TDS models) output is within limits for gain:
a. Measure gain:
H
Move the precision 50
W
cable from the rear-panel DELAYED
TRIGGER OUTPUT BNC to the rear-panel SIGNAL OUT BNC.
H
Push TRIGGER MENU.
Summary of Contents for TDS 500C
Page 4: ......
Page 15: ...Performance Verification Procedures ...
Page 16: ......
Page 115: ...Specifications ...
Page 116: ......
Page 132: ...Nominal Traits 2 16 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Page 147: ......
Page 148: ......