TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
Figure 1–34: 60 Hz Rejection test result – TDS 684B shown
6. Check Line Count Accuracy.
a. Set up oscilloscope for Line Count Accuracy Test:
H
Press WAVEFORM OFF.
H
Press CH1.
H
Press HORIZONTAL MENU.
H
Press the main-menu Record Length.
H
Press the side-menu –more– until you see the appropriate menu.
H
Press the side-menu 5000 points in 100divs.
H
Press the main-menu Horiz Scale (/div).
H
Use the keypad to set the horizontal scale to 200 ns (press 200,
SHIFT, n, then ENTER).
b. Check Line Count Accuracy:
H
Connect a composite output signal from the rear of the PAL signal
source (labeled COMPST on the TSG 121) to the CH1 input
through a 75
W
cable and a 75
W
terminator. See Figure 1–35.