Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–27
Signal Acquisition System Checks
These procedures check those characteristics that relate to the signal-acquisition
system and are listed as checked under Warranted Characteristics in Specifica-
tions.
Equipment
required
None
Prerequisites
The oscilloscope must meet the prerequisites listed on page 1–15.
See Input Channels vs. Model on page 1–2.
1. Preset the instrument controls:
a. Initialize the oscilloscope:
H
Press save/recall SETUP.
H
Press the main-menu button Recall Factory Setup.
H
Press the side-menu button OK Confirm Factory Init.
H
Press CLEAR MENU to remove the menus from the screen.
b. Modify the default settings:
H
Press SHIFT; then press ACQUIRE MENU.
H
On the TDS 600B, press the main-menu button Mode; then press the
side-menu button Average 16.
H
On the TDS 500C and 700C, press the main-menu button Mode;
then press the side-menu button Hi Res.
H
Press CURSOR.
H
Press the main-menu button Function; then press the side-menu
button H Bars.
H
Press CLEAR MENU.
H
Be sure to disconnect any input signals from all four channels.
2. Confirm input channels are within limits for offset accuracy at zero offset:
Do the following substeps — test CH 1 first, skipping substep a since CH 1
is already set up to be checked from step 1.
a. Select an unchecked channel: Press WAVEFORM OFF to remove the
channel just confirmed from the display. Then, press the front-panel
button that corresponds to the channel you are to confirm.
Check Offset Accuracy
(Zero Setting)