Performance Tests
1–74
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
Digitizing Oscilloscope
50
cable
Signal Generator
Output
Figure 1–31: 60 Hz Rejection test hookup
H
Adjust the signal generator for three vertical divisions of 60 Hz
signal. See Figure 1–32. The signal will not be triggered. That is, it
will run free.
Figure 1–32: 60 Hz Rejection test setup signal
c.
Check 60 Hz rejection:
H
Use the keypad to set the horizontal scale (/div) to 50
s (press 50,
SHIFT,
m
, then ENTER).