![Tektronix TDS 500C Performance Verification Manual Download Page 77](http://html1.mh-extra.com/html/tektronix/tds-500c/tds-500c_performance-verification-manual_1079136077.webp)
Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–61
H
Press the main-menu button Source.
H
Press the side-menu button Ch3.
(Ax1 on some TDS models)
H
Set vertical SCALE to 100 mV.
H
Press SET LEVEL TO 50%.
H
Press MEASURE; then press the main-menu button Select
Measrmnt for Ch2.
H
Repeatedly press the side-menu button –more– until Pk-Pk appears
in the side menu (its icon is shown at the left). Press the side-menu
button Pk-Pk.
H
Press CLEAR MENU.
b. Check against limits:
H
CHECK that the readout Ch2 Pk-Pk is between 80 mV and 120 mV,
inclusive, for the TDS 600B or is between 88 mV and 132 mV,
inclusive, for the TDS 500C/700C.
H
Enter voltage on test record.
H
Press VERTICAL MENU; then press the side-menu button
to
toggle to the 50
W
setting.
H
Press CLEAR MENU.
H
CHECK that the readout Ch2 Pk-Pk is between 40 mV and 60 mV,
inclusive, for the TDS 600B or is between 44 mV and 66 mV,
inclusive, for the TDS 500C/700C.
H
Enter voltage on test record.
4. Disconnect the hookup: Disconnect the cables from the channel inputs and
the rear panel outputs.