TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
2–17
Warranted Characteristics
This section lists the various warranted characteristics that describe the
TDS 500C, TDS 600B and TDS 700C oscilloscopes. Electrical and environmen-
tal characteristics are included.
Warranted characteristics are described in terms of quantifiable performance
limits which are warranted.
NOTE
. In these tables, those warranted characteristics that are checked in the
procedure Performance Verification appear in boldface type under the column
Name.
As stated above, this section lists only warranted characteristics. A list of typical
characteristics starts on page 2–25.
Performance Conditions
The performance limits in this specification are valid with these conditions:
H
The oscilloscope must have been calibrated/adjusted at an ambient tempera-
ture b20
_
C and +30
_
C.
H
The oscilloscope must be in an environment with temperature, altitude,
humidity, and vibration within the operating limits described in these
specifications.
H
The oscilloscope must have had a warm-up period of at least 20 minutes.
H
The oscilloscope must have had its signal-path-compensation routine last
executed after at least a 20 minute warm-up period at an ambient temperature
within
±
5
_
C of the current ambient temperature.