Brief Procedures
1–6
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
H
Press SHIFT; then press UTILITY.
H
Highlight Cal in the pop-up menu by repeatedly pressing the
main-menu button System. See Figure 1–2.
H
Verify that the word Pass appears in the main menu under the
following menu labels: Voltage Reference, Frequency Response,
and Pulse Trigger. See Figure 1–2.
f.
Run the signal-path compensation: Press the main-menu button Signal
Path; then press the side-menu button OK Compensate Signal Paths.
g. Wait: Signal-path compensation may take five minutes on the
TDS 500C/700C and fifteen minutes on the 600B to run. While it
progresses, a “clock” icon (shown at left) is displayed on-screen. When
compensation completes, the status message will be updated to Pass or
Fail in the main menu. See step h.
h. Confirm signal-path compensation returns passed status: Verify that the
word Pass appears under Signal Path in the main menu. See Figure 1–2.
Verify Pass
3
Highlight Cal
1
Verify Pass
2
Figure 1–2: Verifying adjustments and signal-path compensation
2. Return to regular service: Press CLEAR MENU to exit the system menus.