![Tektronix TDS 500C Performance Verification Manual Download Page 139](http://html1.mh-extra.com/html/tektronix/tds-500c/tds-500c_performance-verification-manual_1079136139.webp)
Warranted Characteristics
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
2–23
Table 2–14: Warranted characteristics — Environmental (cont.)
Name
Description
Susceptibility
1, 2
Meets or exceeds the EMC requirements of the following standards:
EN 50082–1
European Community Requirements
IEC 801-2
Electrostatic Discharge Performance Criteria B
IEC 801-3
Radiated Susceptibility 3 V/meter from 27 MHz to 500 MHz
unmodulated
IEC 801-4
Fast Transients Performance Criteria B
IEC 801-5
AC Surge Performance Criteria B
Approvals
Conforms to and is certified where appropriate to:
UL 3111–1
3
– Standard for electrical measuring and test equipment
CAN/CSA C22.2 no. 1010.1
3
– Safety requirements for electrical equipment for
measurement, control and laboratory use
1
VGA output cable needs to be terminated, if connected at all, for the Instrument to meet these standards. The test will
pass with LCOM part # CTL3VGAMM–5.
2
The GPIB cable connected to the instrument for certain of the emissions tests must be “low EMI” having a high-quality
outer shield connected through a low impedance to both connector housings. Acceptable cables are Tektronix part
numbers 012-0991-00, -01, -02, and -03. In order to maintain the EMI performance conforming to the above regulations,
the following cables, or their equivalent, should be used: a shielded Centronics cable, 3 meters in length, part number
012-1214-00, and a shielded RS-232 cable, 2.7 meters in length, CA part number 0294-9.
3
UL 3111, CSA 22.2 no.1010 Safety Certification Compliance:
Temperature (operating) 5 to +40 C
Altitude (maximum operating): 2000 meters
Equipment Type: Test and Measurement
Safety Class: Class I (as defined in IEC 1010–1, Annex H) – grounded product
Overvoltage Category: Overvoltage Category II (as defined in IEC 1010–1, Annex J)
Pollution Degree: Pollution Degree 2 (as defined in IEC 1010–1)
Note – Rated for indoor use only