![Tektronix TDS 500C Performance Verification Manual Download Page 27](http://html1.mh-extra.com/html/tektronix/tds-500c/tds-500c_performance-verification-manual_1079136027.webp)
Brief Procedures
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–11
c.
The horizontal POSITION knob positions the signal left and right
on-screen when rotated.
3. Remove the test hookup: Disconnect the probe from the channel input and
the probe-compensation terminals.
Equipment
required
One probe such as the P6243, P6245, or P6139A
Prerequisites
None
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel;
connect the probe ground to PROBE COMPENSATION GND. See
Figure 1–3 on page 1–8.
b. Initialize the oscilloscope:
H
Press save/recall SETUP.
H
Press the main-menu button Recall Factory Setup.
H
Press the side-menu button OK Confirm Factory Init.
c.
Modify default settings:
H
Set the vertical SCALE to 200 mV.
H
Set the horizontal SCALE for the M (main) time base to 200
s.
H
Press SET LEVEL TO 50%.
H
Press TRIGGER MENU.
H
Press the main-menu button Mode & Holdoff.
H
Press the side-menu button Normal.
H
Press CLEAR MENU to remove the menus from the screen.
2. Verify that the main trigger system operates: Confirm that the following
statements are true.
H
The trigger level readout for the main trigger system changes with the
trigger-LEVEL knob.
H
The trigger-LEVEL knob can trigger and untrigger the square-wave
signal as you rotate it. (Leave the signal untriggered, which is indicated
by the display not updating).
Verify the Main and
Delayed Trigger Systems
Summary of Contents for TDS 500C
Page 4: ......
Page 15: ...Performance Verification Procedures ...
Page 16: ......
Page 115: ...Specifications ...
Page 116: ......
Page 132: ...Nominal Traits 2 16 TDS 500C TDS 600B TDS 700C Performance Verification and Specifications ...
Page 147: ......
Page 148: ......