![Tektronix TDS 500C Performance Verification Manual Download Page 50](http://html1.mh-extra.com/html/tektronix/tds-500c/tds-500c_performance-verification-manual_1079136050.webp)
Performance Tests
1–34
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
NOTE
. Refer to the Sine Wave Generator Leveling Procedure on page 1–84 if
your sine wave generator does not have automatic output amplitude leveling.
c.
Hook up the test-signal source: Connect the sine wave output of a
leveled sine wave generator to CH 1. Set the output of the generator to a
reference frequency of 10 MHz or less. See Figure 1–7. For the optional
setup using a leveled sine wave generator with a leveling head (item 26)
see Figure 1–8 and, if using this optional setup with the example
Tektronix SG 504, set the generator output to 6 MHz.
Digitizing Oscilloscope
Sine Wave
Generator
Output
Figure 1–7: Initial test hookup
High Frequency Sine
Wave Generator
(example: Tek SG 504)
Digitizing Oscilloscope
Output
Leveling head
Figure 1–8: Optional initial test hookup
2. Confirm the input channels are within limits for analog bandwidth: Do the
following substeps — test CH 1 first, skipping substeps a and b since CH 1
is already set up for testing from step 1.
a. Select an unchecked channel:
H
Press WAVEFORM OFF to remove the channel just confirmed
from display.