RUN EXCITE-2008: Service & Maintenance Manual - rev. 1.3
Page 10.1
10.
FUNCTIONAL TEST MD MODELS
Every time maintenance and/or repairs are performed on a machine, it is necessary to carry out a
series of mechanical and electrical tests to ensure:
•
that the medical device is working properly;
•
that the medical device is working safely;
•
that the exercises are performed according to the biomedical specifications for which the
medical device was designed.
Technogym ensures its own safety standards by means of the following tools and activities:
•
ISO 9001/2000 certification of the company;
•
certification of the device in compliance with medical regulations 60601-1 and 60601-1-2;
•
electrical testing of each device manufactured, according to the provisions of the
aforementioned standards;
•
training of technical staff who perform technical service;
•
use of checked and tested original Technogym spare parts for all service operations;
10.1.
ELECTRICAL SAFETY TEST
Based on the risk analysis carried out by the After Sale department regarding installation and
technical service, it is still necessary to use a tester to perform a safety test on the grounding ring.
This test must ensure a continuous earth connection (resistance less than 4 Ohm) between the
grounding pole of the mains lead (connected to the machine) and the 6 points listed below:
•
screw fastening the display rear casing to the frame
•
screw fastening the uprights to the frame;
•
screw fastening the rear casing to the frame;
For more complete explanations, see the boxes in the figure below.
Summary of Contents for EXCITE Run 500
Page 1: ...D446 D448 D449 SERVICE MAINTENANCE MANUAL REV 1 3...
Page 2: ......
Page 4: ......
Page 32: ...RUN EXCITE 2008 Service Maintenance Manual rev 1 3 Page 2 20 Page intentionally left blank...
Page 52: ...RUN EXCITE 2008 Service Maintenance Manual rev 1 3 Page 3 20 Page intentionally left blank...
Page 59: ...RUN EXCITE 2008 Service Maintenance Manual rev 1 3 Page 4 7 Page intentionally left blank...
Page 60: ......
Page 127: ...RUN EXCITE 2008 Service Maintenance Manual rev 1 3 Page 7 19 Continued on following page...
Page 148: ......
Page 217: ...RUN EXCITE 2008 Service Maintenance Manual rev 1 3 Page 9 63 Page intentionally left blank...
Page 218: ......
Page 229: ...RUN EXCITE 2008 Service Maintenance Manual rev 1 3 Page 11 7 Page intentionally left blank...
Page 230: ......
Page 233: ......