13. Trouble Shooting
13-2
Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1
2004-10
Error number
Error text
Possible Defective Part
See chapter…
0x0303
"Erasing flash chip
failed"
Problem with the main board
8.5Replacing the Main
Board
0x0304
"Erasing flash
block failed"
Problem with the main board
8.5Replacing the Main
Board
0x0305
"Unexpected flash
device"
Problem with the main board
8.5Replacing the Main
Board
0x0306
"Flash is busy"
Problem with the main board
8.5Replacing the Main
Board
0x0401
"Filter is busy"
Filterslide can not be inserted during
measurement
---
0x0402
"Already inserted"
Filterslide is already inserted
---
0x0403
"No filter carriage
detected"
Check filterslide coding slits
---
0x0404
"No measurement
filter defined"
Check filter settings
4.4.8 User Configuration
0x0405
"No reference filter
defined"
Check filter settings
4.4.8 User Configuration
0x0407
"Wavelength %d
nm not available"
Filter value out of range
---
0x0408
"Filter carriage not
defined =>
Type:%d,
Number:%d"
Check filterslide coding slits
---
0x0409
"No Gradient Filter
detected"
Check filterslide coding slits
---
0x0601
"Wrong checksum
in configuration
block 1"
Check filter settings
4.4.8 User Configuration
0x0602
"Wrong checksum
in all configuration
blocks !"
Check filter settings
4.4.8 User Configuration
0x0603
"Writing of all
configuration
blocks failed !"
Check filter settings
4.4.8 User Configuration,
4.4.3 Setup All
0x0701
"No data
available"
Error is displayed when kinetic
measurements got aborted
(instrument is set to Spectra Mode)
----
0x0702
"No data available
for measurement
cycle no. %d"
Error is displayed when kinetic
measurements got aborted
(instrument is set to Spectra Mode)
----
0x0804
"Area 340 not
adjusted"
Problem with the main board
4.4.5 Mirror and lamp
adjustment
0x0805
"Area 400 not
adjusted"
Problem with the main board
4.4.5 Mirror and lamp
adjustment
0x0806
"E2Pot Overflow"
Problem with the main board
4.4.5 Mirror and lamp
adjustment
0x0901
"Illegal Baudrate:
%d"
Check instrument port settings
4.3.3 Define Instrument
Mode
0x0902
"Illegal
DataLength: %d"
Check instrument port settings
4.3.3 Define Instrument
Mode
0x0903
"Illegal # of
StopBits: %d"
Check instrument port settings
4.3.3 Define Instrument
Mode
0x0904
"Illegal Parity: %d"
Check instrument port settings
4.3.3 Define Instrument
Mode