4. Firmware and Software Description
4-8
Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1
2004-10
4.4 SunSet Software (Service tool kit is necessary!)
This program enables the user to adjust the instrument:
•
Device configuration
•
Lamp adjust
•
Plate calibration
•
Measurement offsets
•
User configuration
4.4.1
Installation of SUNSET Software
The SunSet Software is installed using the following procedure:
•
Double click the setup icon to automatically install the SunSet software or
click the
Start
button on the lower taskbar and select <Run…> from the
menu and then browse for the correct setup file.
•
A series of dialog boxes will appear, read each one, enter any necessary
information and click
Next
to continue.
•
The files are then installed and the program icon is created.
•
When the
Installation
Complete
dialog box appears, click
Finish
and
SunSet Software is operational.
4.4.2
Starting the SunSet Software
Note:
•
Perform the Sunset test with filter slide G or filter slide A.
•
Filterslide G ensures that the entire
wavelength range is checked (340 - 750 nm)
•
It is not possible to do the adjustments
with the Tuneable Wavelength filterslide.
If an instrument is already connected to one of Tecan’s programs, close the
program or disconnect the instrument from the program.
To start the SunSet Software click the
Start
button on the lower taskbar and
select
Programs
and then
TECAN
–from the menus and select
SunSet
or click
the SunSet icon in the directory C:\Programme\Tecan\SunSet.
The following dialog box is displayed:
Select
Connect
from the Setup menu or press <Strg> and <F5> together to
connect the instrument.