12. Performance Testing / Quality Control
12-2
Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1
2004-10
12.2.2 Operating
Procedure
General
1. The best repeatability is obtained, when the measurement wavelength
corresponds to the maximum absorbance wavelength of the particular
solution.
It is important to use the maximum absorbance wavelength, if the
absorbance curve of the sample is over a narrow wavelength band.
2. After each microplate has been measured, please refer to the test kit
package for information regarding the validation procedure.
3. When very accurate results are required, ensure the
Accurate
measurement
mode is used.
Microplates
1. The instrument can be used with those types of microplates which are
described in the specifications in chapter 2.3.8