4.3.4 Define
Filter
The following dialog box is displayed:
Click the Filter slide
Out
button to move the filter out of the instrument.
To insert a filter slide, open the filter compartment manually and slide the filter
into the slot, so that the filter end of the slide is inserted first. (Do not force the
filter slide into the instrument beyond the point of resistance).
Click the Filter slide
In
button and the filter is inserted.
Pos1 - 4
show the filter values for the currently loaded absorbance filters.
Note:
The instrument is able to recognize predefined filter slides and the filter
values for these slides must not be changed. However, if the filters in the
filter slide have been changed (by a service engineer) or if a new
undefined customized filter slide is to be used, the filter slides need to be
defined.
To define the filter values for a new filter slide, enter the required wavelengths in
the text boxes. Click
Next
.
Note:
The wavelength range for the
sunrise
is 340 - 750 nm.
It is not possible to define the tuneable wavelength filter slide from the
Sunset Instrument Settings