Useful Function
57
Linearity Compensation
This function provides a high output setting accuracy by measuring and compensating the nonlinearity of the output
caused by feedback systems or DA converters/Error Amplifiers used as a DC power supply voltage or current
reference value.
This is normally used when the compensation function is turned ON (Parameter 1 or 2).
Setting Procedures
Press The Function “FUNC.” key.
The FUNC. lamp will light.
The Setting Item Number will be displayed in the Number
Display (voltage) and the parameter will be displayed in
the Number Display (current).
Select Item number 12 with the VOLTAGE dial and
Select the operation mode of compensation with the CURRENT dial.
0 = Compensation Function is turned OFF
1 = Once compensation has concluded, the Output is then
turned ON, the set values are changed, and the
compensation value is maintained until there is a change in
the Output mode (CV/CC).
2 = Compensates continually (200ms cycles)
* The default setting is "1".
Press the FUNC. key to finish.
The "FUNC." lamp is turned off and the Number Display
returns to the measurement display mode.
Summary of Contents for HX-S-G4 Series
Page 157: ...Special Loads 157 This page is blank...
Page 161: ...Maintenance 161 This page is blank...
Page 197: ...Specifications 197 External Diagrams 1 HX S 030 200G4 F I...
Page 198: ...Specifications 198 2 HX S 030 400G4 F I...
Page 199: ...Specifications 199 3 HX S 060 100G4 F I...
Page 200: ...Specifications 200 4 HX S 060 200G4 F I...
Page 201: ...Specifications 201 5 HX S 0500 12G4 F I...
Page 202: ...Specifications 202 6 HX S 0500 24G4 F I...
Page 203: ...Specifications 203 7 HX S 01000 6G4 F I...