Calibration
183
7. Set the constant voltage setting value to [ xV ] and the
constant current setting value to [ yA ] and set the
output to ON.
30-200
30-400
60-100
60-200
[xV]
30.00V
30.00V
60.00V
60.00V
[yA]
1.0A
1.0A
1.0A
1.0A
500-12
500-24
1000-6
1000-12
[xV]
500.0V
500.0V
1000V
1000V
[yA]
0.10A
0.10A
0.010A
0.10A
On how to set, see Operating as Constant Voltage
Power Supply under Basic Operations.(page27)
8. Press the FUNC. key.
9. Turn the VOLTAGE dial to display the setting item
number 25 in the Number Display (voltage).
* At that time, the voltage measured value is displayed in
the Number Display (current).
10. Adjust the CURRENT dial so that the voltage
measured value displayed in the Number Display
(current) is the same as the actual output voltage
value (digital multimeter measurement).
* The figure above is an example of connecting (HX030-200G4)
11. Press the FUNC. key to settle the full-scale
calibration value in the voltmeter.
12. This completes offset/full-scale calibration of
voltmeter.
Set the constant voltage setting value and the
constant current setting value to the original
ones and start using.
Summary of Contents for HX-S-G4 Series
Page 157: ...Special Loads 157 This page is blank...
Page 161: ...Maintenance 161 This page is blank...
Page 197: ...Specifications 197 External Diagrams 1 HX S 030 200G4 F I...
Page 198: ...Specifications 198 2 HX S 030 400G4 F I...
Page 199: ...Specifications 199 3 HX S 060 100G4 F I...
Page 200: ...Specifications 200 4 HX S 060 200G4 F I...
Page 201: ...Specifications 201 5 HX S 0500 12G4 F I...
Page 202: ...Specifications 202 6 HX S 0500 24G4 F I...
Page 203: ...Specifications 203 7 HX S 01000 6G4 F I...