Calibration
171
7. Set the external resistance to 10kΩ and the
constant current setting value to [ yA ] and set
the output to ON.
30-200
30-400
60-100
60-200
[yA]
1.0A
1.0A
1.0A
1.0A
500-12
500-24
1000-6
1000-12
[yA]
0.10A
0.10A
0.010A
0.10A
On how to set, refer to Output Voltage Control with
External Resistance (B).(page139)
8. Press the FUNC. key.
9. Turn the VOLTAGE dial to display the setting item
number 29 in the Number Display (voltage).
* At that time, "C" or "F" is displayed in the Number
Display (current).
C: Coarse adjustment
F: Fine adjustment
10. Adjust so that the output voltage becomes [ xV ]
with the CURRENT dial.
30-200
30-400
60-100
60-200
[xV]
30.000V
30.000V
60.000V
60.000V
500-12
500-24
1000-6
1000-12
[xV]
500.00V
500.00V
1000.0V
1000.0V
* Pressing the CURRENT dial switches the
adjustment between "C" (coarse adjustment) and
"F" (fine adjustment).
11. Press the FUNC. key to settle the full-scale
calibration value of output voltage control with
external resistance (B).
12. This completes offset/full-scale calibration of
output voltage control with external resistance (B).
Summary of Contents for HX-S-G4 Series
Page 157: ...Special Loads 157 This page is blank...
Page 161: ...Maintenance 161 This page is blank...
Page 197: ...Specifications 197 External Diagrams 1 HX S 030 200G4 F I...
Page 198: ...Specifications 198 2 HX S 030 400G4 F I...
Page 199: ...Specifications 199 3 HX S 060 100G4 F I...
Page 200: ...Specifications 200 4 HX S 060 200G4 F I...
Page 201: ...Specifications 201 5 HX S 0500 12G4 F I...
Page 202: ...Specifications 202 6 HX S 0500 24G4 F I...
Page 203: ...Specifications 203 7 HX S 01000 6G4 F I...