Calibration
193
7. Set the constant voltage setting value to [ xV ] and
the constant current setting value to [ yA ] and set
the output to ON.
30-200
30-400
60-100
60-200
[xV]
1.00V
1.00V
1.00V
1.00V
[yA]
200.0A
400.0A
100.0A
200.0A
500-12
500-24
1000-6
1000-12
[xV]
1.0V
1.0V
1V
1V
[yA]
12.00A
24.00A
6.000A
12.00A
120-50
120-100
400-15
400-30
[xV]
1.0V
1.0V
1.0V
1.0V
[yA]
50.00A
100.0A
15.00A
30.00A
See Operating as Constant Current Power Supply
under Basic Operations on how to set. (page
8. Press the FUNC. key.
9. Turn the VOLTAGE dial to display the setting item
number 27 in the Number Display (voltage).
* At that time, the current measured value is displayed
in the Number Display (current).
* The figure above is an example of connecting (HX-S-030-
200G2)
10. Adjust the CURRENT dial so that the current
measured value displayed in the Number Display
(current) is the same as the actual output current
value (digital multimeter measurement).
* The figure above is an example of connecting (HX-S-030-
200G2)
11. Press the FUNC. key to settle the full-scale
calibration value in the ammeter.
12. This completes offset/full-scale calibration of
ammeter.
Set the constant voltage setting value and the
constant current setting value to the original ones
and start using.
Summary of Contents for HX-S-01000-12G2
Page 165: ...Special Loads 165 This page is blank...
Page 169: ...Maintenance 169 This page is blank...
Page 211: ...Specifications 211 External Diagrams 1 HX S 030 200G2 F I...
Page 212: ...Specifications 212 2 HX S 030 400G2 F I...
Page 213: ...Specifications 213 3 HX S 060 100G2 F I...
Page 214: ...Specifications 214 4 HX S 060 200G2 F I...
Page 215: ...Specifications 215 5 HX S 0120 50G2 F I...
Page 216: ...Specifications 216 6 HX S 0120 100G2 F I...
Page 217: ...Specifications 217 7 HX S 0400 15G2 F I...
Page 218: ...Specifications 218 8 HX S 0400 30G2 F I...
Page 219: ...Specifications 219 9 HX S 0500 12G2 F I...
Page 220: ...Specifications 220 10 HX S 0500 24G2 F I...
Page 221: ...Specifications 221 11 HX S 01000 6G2 F I...
Page 222: ...Specifications 222 12 HX S 01000 12G2 F I...
Page 223: ...Specifications 223...
Page 239: ......