Calibration
185
7. Set the constant voltage setting value to [ xV ] and
the external resistance to 0Ω and set the output to
ON.
30-200
30-400
60-100
60-200
[xV]
1.00V
1.00V
1.00V
1.00V
500-12
500-24
1000-6
1000-12
[xV]
1.0V
1.0V
1V
1V
120-50
120-100
400-15
400-30
[xV]
1.0V
1.0V
1.0V
1.0V
On how to set, refer to Output Current Control with
External Resistance (A).(page
8. Press the FUNC. key.
9. Turn the VOLTAGE dial to display the setting item
number 31 in the Number Display (voltage).
* At that time, "C" or "F" is displayed in the Number
Display (current).
C: Coarse adjustment
F: Fine adjustment
10. Adjust so that the output current becomes [ yA ]
with the CURRENT dial.
30-200
30-400
60-100
60-200
[yA]
200.00A
400.00A
100.00A
200.00A
DMM display
value
50.00mV
40.00mV
25.00mV
50.00mV
500-12
500-24
1000-6
1000-12
[yA]
12.000A
24.000A
6.000A
12.000A
DMM display
value
40.00mV
40.00mV
30.00mV
40.00mV
120-50
120-100
400-15
400-30
[yA]
50.000A
100.00A
15.000A
30.000A
DMM display
value
12.50mV
25.00mV
50.00mV
50.00mV
※
*DMM display value is the specification of the shunt resistor
The value when had you use.
* Pressing the CURRENT dial switches the
adjustment between "C" (coarse adjustment) and
"F" (fine adjustment).
11. Press the FUNC. key to settle the full-scale
calibration value of output current control (A)
with external resistance.
12. This completes offset/full-scale calibration of
output current control with external resistance (A).
Set the constant voltage setting value and the
constant current setting value to the original ones
and start using.
Summary of Contents for HX-S-01000-12G2
Page 165: ...Special Loads 165 This page is blank...
Page 169: ...Maintenance 169 This page is blank...
Page 211: ...Specifications 211 External Diagrams 1 HX S 030 200G2 F I...
Page 212: ...Specifications 212 2 HX S 030 400G2 F I...
Page 213: ...Specifications 213 3 HX S 060 100G2 F I...
Page 214: ...Specifications 214 4 HX S 060 200G2 F I...
Page 215: ...Specifications 215 5 HX S 0120 50G2 F I...
Page 216: ...Specifications 216 6 HX S 0120 100G2 F I...
Page 217: ...Specifications 217 7 HX S 0400 15G2 F I...
Page 218: ...Specifications 218 8 HX S 0400 30G2 F I...
Page 219: ...Specifications 219 9 HX S 0500 12G2 F I...
Page 220: ...Specifications 220 10 HX S 0500 24G2 F I...
Page 221: ...Specifications 221 11 HX S 01000 6G2 F I...
Page 222: ...Specifications 222 12 HX S 01000 12G2 F I...
Page 223: ...Specifications 223...
Page 239: ......