Figure 10: Quick Coupling for Kirschner Wires
B
0.6 to 3.2 mm for UNIUM™
(05.001.188)
Figure 4
Figure 9
Figure 6
Figure 7
Figure 8
3. Spray and Wipe
Spray the device with newly prepared neutral pH enzy-
matic or mild al kaline detergent solution in accordance to
the detergent manufacturer’s instructions. The tempera-
ture of the solution should be ≤ 40 °C (≤ 104 °F)
for manual cleaning. Leave the solution to interact with
the surface of the device for a minimum of 2 minutes
(Figs. 4, 5 and 6) and then wipe it off.
4. Rinse with Tap Water
Rinse device with cold tap water for a minimum of
2 minutes (Fig. 7). Use a syringe or pipette to flush lumens
and channels.
5. Clean with Detergent
Manually clean device under running warm water for a
minimum of 5 minutes. From time to time (during the
5 minutes cleaning period) use a soft-bristled brush
and/or soft lint-free cloth that is soaked with the newly
prepared neutral pH enzymatic or mild alkaline deter-
gent solution to remove soil and debris (Fig. 8 and
Fig.9). Manipulate all moving parts such as triggers,
sliding sleeves, attachment release rings, saw blade
coupling, switches, etc. under running water (Fig. 10).
Care and Maintenance
Cleaning and Disinfection
Manual Cleaning Instructions
Figure 5
31
UNIUM
™
• Instruction for Use