Figure 3
Oscillating Saw Attachment (532.021),
17,000 osc./min
Positioning the Oscillating Saw Attachment
The oscillating saw attachment can be inserted in eight
different positions (45° increments).
Slide the sleeve (1) toward the attachment coupling and
rotate it into the desired position (Fig. 1).
▲ Precaution:
To prevent injury, always hold the saw attachment with
the saw blade facing away from the user.
Changing the Saw Blade
1. Pull the locking knob (2) down and turn it
counterclockwise.
2. Lift and remove the saw blade.
3. Insert a new saw blade into the saw blade coupling (3)
and move it into the desired position (Fig. 2).
The desired positions can be offset from each other at
45º angles (Fig. 3).
4. Place your thumb on the saw blade coupling (3) to hold
the saw blade and turn the locking knob (2) clockwise
until the saw blade is secured.
5. Put the mode switch of the handpiece in the ON or
FWD ONLY position.
▲ Precautions:
Saw blades labeled “Single Use” should not be reused.
Using the Oscillating Saw Attachment
The saw blade must already be oscillating when the saw
is applied to the bone. Do not apply strong pressure to
the saw blade as this will delay the cutting process and
the saw teeth will catch in the bone. Optimal saw perfor-
mance is achieved by moving the Power Tool slightly
back and forth in the plane of the saw blade so that the
blade oscillates beyond the bone on both sides. Very
precise cuts can be made when the saw blade is guided
steadily. Imprecise cuts arise due to used saw blades,
excess pressure or the saw blade jamming on the bone.
Attachments
Saw Attachments
Fine tooth saw blades
Coarse tooth saw blades
Parallel saw blades
2
3
1
Figure 1
Figure 2
22