2
1
70 mm
Oscillating Saw Attachment II
(Crescentic Technique) (532.023), 17,000 osc./min
The Oscillating Saw Attachment II is designed for use
with crescentic saw blades guided by a 1.6 mm Kirschner
Wire. It can also be used with saw blades that have a
shaft extension of 105° for reaching difficult- to- access
sites.
Inserting and Removing the Saw Blades
Pull the sliding sleeve (2) back and then insert the saw
blade until it locks into the saw blade coupling (1). Release
the sliding sleeve (2) and check that the saw blade is
correctly fixed by gently pulling on it.
Pull the sliding sleeve (2) back and remove the saw blade.
Figure 2: Crescentic Saw Blades e.g. (03.000.313S)
Figure 3: 105° Angled Saw Blades e.g. (03.000.341S)
Figure 1: Oscillating Saw Attachment II (Crescentic Technique)
Attachments
Saw Attachments
23
UNIUM
™
• Instruction for Use