6. Rinse with Tap Water
Rinse the device thoroughly using cool to lukewarm run-
ning water for a minimum of 2 minutes (Fig. 11). Use a sy-
ringe, pipette or water jet to flush lumens and channels.
Move joints, handles and other movable device features
in order to rinse thoroughly under running water. After
complet ing this step, drain any excess water.
7. Wipe/Spray Disinfection
Wipe off or spray the surfaces of the devices with a
minimum of 70% alcohol-based disinfectant for thirty
seconds. This step has to be repeated two (2) additional
times using a new, clean, soft and lint-free cloth damp-
ened with a minimum 70% alcohol-based disinfectant
each time.
8. Visually Inspect Device
Inspect the device including cannulations, sliding sleeves,
attachment release rings, etc. for visible soil. Repeat the
process if visible soil remains.
9. Final Rinse with De-ionized/Purified
Water
Final rinse with de-ionized or purified water for a mini-
mum of 2 minutes (Fig. 12).
10. Dry
Dry device using a clean, soft lint-free cloth (Fig. 13) or
medical grade compressed air. If smaller devices or can-
nulations contain residual water, blow them dry with
medical grade compressed air. Ensure that all parts are
dry including the inside of the handpiece.
Figure 13
Figure 12
Figure 11
Care and Maintenance
Cleaning and Disinfection
Manual Cleaning Instructions
32
Instruction for Use • UNIUM
™