UNIUM™ Handpieces
UNIUM™ Modular Handpiece
(05.001.601)
1 Mode Switch: ON (forward, reverse, oscillating drilling
modes), OFF (lock mode), FWD ONLY (forward only
mode)
2 Forward Trigger
3 Reverse Trigger
4 Attachment Coupling
5 Release Buttons for Attachment
6 Casing Opening Button on the Front Lid
7 Casing Cover with Battery Level Indicator
Forward Mode
Set the Mode Switch (1) in the ON or FWD ONLY position
and press the Forward Trigger (2) to gradually increase
and decrease the speed.
Reverse Mode
Set the Mode Switch (1) in the ON position and press the
Reverse Trigger (3) to gradually increase and decrease the
speed.
Oscillating Drilling Mode
Set the Mode Switch (1) in the ON position and press the
Forward and Reverse Triggers (2 and 3) at the same time.
To gradually increase the oscillating drilling speed, press
down both triggers simultaneously. To decrease the
speed, simultaneously reduce the pressure on both trig-
gers. The oscillating drilling mode helps protect against
soft tissue damage when drilling and inserting Kirschner
wires. The cutting tool inserted into the attachment
oscillates both clockwise and counterclockwise.
Both triggers (2) and (3) allow control of the speed from
0 to the maximum rpm. Maximum torque and speed
vary depending on the attachment. Ensure that the
correct attachment is used for each operation in terms
of speed and torque.
▲ WARNINGS:
•
Only use the oscillating drilling mode with drill
attachments (05.001.250, 05.001.252, 05.001.253,
532.011, 532.012) or the K-wire attachment
(05.001.188). Never use the oscillating drilling mode
with any other attachment to avoid injuries.
•
Never use the saw attachments (532.021, 532.023 and
532.026) in oscillating drilling mode to avoid injuries.
Modular Handpiece
4
11
UNIUM
™
• Instruction for Use