Radiolucent Drive (511.300, 511.30*), 1,255 rpm
The Radiolucent Drive can be used with the UNIUM
Modular Handpiece in combination with the
AO/ASIF Quick Coupling (05.001.250) and the Adapter
for the Radiolucent Drive (532.031).
Coupling the Radiolucent Drive
Connect the AO/ASIF Quick Coupling to the UNIUM
Modular Handpiece and the adapter to the Quick
Coupling. Position the Radiolucent Drive as far as it will go
over the Quick Coupling and the adapter and rotate it into
the desired working position. Support
the drive with your free hand.
Inserting the Drill Bit
To insert the drill bit, push the release sleeve ring (1) on
the attachment and position the drill bit inside the
coupling as far as it can go while rotating it slightly. Pull
back the ring on the attachment to fix the drill. Check if
the drill bit is seated correctly by gently pulling on it
(Fig. 2).
Removing the Drill Bit
Follow the same procedure in reverse order.
Using the Radiolucent Drive
Before positioning the Radiolucent Drive, align the image
intensifier until the distal locking hole of the medullary nail
is round and easily visible (Fig. 3).
After the incision, position the Radiolucent Drive and
center the drill bit tip over the locking hole. On the
monitor of the image intensifier, you can see both the drill
bit and the target rings of the drive (Fig. 4).
Swing the drive up and center it precisely so that the drill
bit appears as a round point and the locking hole is visible
around it. The target ring also assists the centering. The
locking hole can now be drilled directly (Fig. 5).
Attachments
Radiolucent Drive
Figure 3
Figure 4
Figure 5
Figure 2
Figure 1
* available in USA only