click
4. The scrubbed person closes the casing cover from the
outside without contacting the Power Unit or the inside
of the casing (Fig. 8–11). The scrubbed person releases
the “front lid” by pressing on the “Press” button (Fig. 8).
The scrubbed person places one hand under the cas-
ing cover (Fig. 9) and closes it. They then turn the “front
lid” 180° until it clicks (Fig. 10). Always ensure that the
casing cover and front lid are totally closed before
using the handpiece (Fig. 11).
▲ WARNING:
The scrubbed person must not touch the non-sterile
Power Unit or the inside of the handpiece in order to
avoid contamination. Should this happen, the scrubbed
nurse must be scrubbed again. Should the handpiece be
contaminated, it must be cleaned and resterilized before
being used in the operating room.
Removing the Power Unit
1. The scrubbed person opens the casing as described in
step 1. The circulating person removes the Power Unit
by holding the handle while the scrubbed person holds
the handpiece firmly (Fig. 12). Ensure that the Power Unit
does not touch the outer surface of the handpiece to
avoid contaminating the Power Unit.
If this occurs follow the information in the section
“Cleaning and Disinfection of the Power Unit
(05.001.602) Directly After the Surgery”. Store the
Power Unit by using the Adapter for UBC II (05.001.604)
in the Universal Battery Charger II (05.001.204) when
not in use. The Universal Battery Charger II should
always be turned on (Fig. 13).
▲ WARNINGS:
•
To ensure aseptic conditions, the Power Unit must not
be removed from the handpiece until the end of
surgery.
•
To avoid injuries, always switch OFF the handpiece
before every manipulation (e.g. inserting/ removing the
power unit, attachments and/or cutting tools) and
before placing it back down.
Figure 9
Figure 8
Figure 11
Figure 10
UNIUM Power Unit
Aseptic Transfer Technique
Figure 12
Figure 13
click
16
Instruction for Use • UNIUM
™