UNIUM™ Power Unit
UNIUM Power Unit (05.001.602)
1 Battery Level Indicator
2 Service Indicator
3 Information Button
4 Handle to hold the Power Unit
The Power Unit (05.001.602) incorporates the Li-Ion
battery and the electronic control unit.
Checking the Battery Level
By briefly pressing the Information Button (3), the battery
level is shown by the Battery Level Indicator (1).
Power Unit Condition Self-Test
A self-test of the Power Unit can be performed by
pressing the Information Button (3) for 8 seconds.
The four battery level indicator LEDs (1) will flash and
the red LED of the Service Indicator (2) will light up.
The self-test is completed after a few seconds. If all the
LEDs turn off, then the Power Unit is in good condition.
If the red LED of the Service Indicator (2) remains on
alone for around 5 seconds, the Power Unit is blocked for
further use and needs to be replaced. The red LED will
then turn off. However, if the Information Button (3) is
pressed down again, the red LED will light up again for a
few seconds.
Charging and Storage of the UNIUM Power Unit with
the Universal Battery Charger II (05.001.204)
In order for the Power Unit to be recognized and charged
by the Universal Battery Charger II (UBC II), a minimum of
firmware 16.0 is required. A label placed on the underside
of the charger indicates which firmware it has. If neces-
sary, send the charger to a DePuy Synthes Representative
for a firmware update.
The UBC II includes four independent charging bays.
Insert the UNIUM Adapter for UBC II (05.001.604) into
one of the bays and the Power Unit into the adapter as
shown in the picture. Up to four Power Units can be
charged simultaneously in separate bays, each equipped
with an adapter.
Store the Power Unit in the UBC II and turn the charging
station on. This will prevent the Power Unit from discharg-
ing and ensures it will be fully charged and ready to use.
To ensure that both UNIUM Handpieces can operate
safely and reliably, the Power Unit has to be checked at
periodical intervals. The UBC II will indicate when a check
is necessary. After the check has been completed the
UBC II will indicate if the Power Unit performance is suffi-
cient or if it needs to be replaced.
For further information on the UBC II and the UNIUM
Adapter for UBC II, consult the Instructions for Use of
UBC II or contact your local DePuy Synthes Office.
Example label found on the underside of the charger with firmware 16.0
SW-Rev. 16.0
2019/10/16
4
2
1
3
Battery Level Indicator (1):
– All four LEDs illuminate when fully charged.
– Three or fewer LEDs illuminate when not fully charged. The state of charge may
suffice depending on the intended application.
– Bottom LED flashes orange when discharged.
13
UNIUM
™
• Instruction for Use